B800-00-880 Issue B<br />

3.7.5 Comm<strong>and</strong> set<br />

Table 16 shows a summary of the full set of comm<strong>and</strong>s available for controlling <strong>and</strong> monitoring the <strong>nEXT</strong> pump.<br />

Table 17 shows the abbreviations that are used to define comm<strong>and</strong>s in the following sections <strong>and</strong> Table 18 shows the<br />

error codes that might be returned.<br />

Table 16 - Summary of comm<strong>and</strong>s that can be sent to the <strong>nEXT</strong> pump<br />

Object name Comm<strong>and</strong> Parameter<br />

range<br />

Node !S850<br />

?S850<br />

Pump type ?S851 7;<br />

10;<br />

4<br />

Pump control !C852 0<br />

1<br />

?V852 0.1800;<br />

32-bits<br />

Vent options !S853<br />

?S853<br />

Timer setting !S854<br />

?S854<br />

Power limit setting !S855<br />

?S855<br />

Normal speed<br />

setting<br />

St<strong>and</strong>by speed<br />

setting<br />

Temperature<br />

readings<br />

Link parameter<br />

readings<br />

!S856<br />

?S856<br />

!S857<br />

?S857<br />

Factory<br />

setting<br />

Data<br />

type<br />

Units<br />

Comments<br />

0.99 - decimal address Multi-drop address<br />

0 = disable multi-drop address<br />

99 = wildcard<br />

0<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

?V859 0.100;<br />

0.100<br />

?V860 0.500;<br />

0.300;<br />

0.15000<br />

- string<br />

string<br />

string<br />

chars<br />

chars<br />

chars<br />

Pump type<br />

DSP software version number<br />

(D396496XXX for <strong>nEXT</strong>)<br />

Full speed RPS (1000 for <strong>nEXT</strong>)<br />

- decimal - Stop the pump<br />

Start the pump<br />

- decimal<br />

hex<br />

RPS<br />

flags<br />

Measured motor speed<br />

System status word<br />

0 decimal Hard vent only when 50% speed or<br />

hard vent if