18<br />
Technical Information<br />
Use<br />
and finally the pump status (for example):<br />
P U M P R E A D Y : U S E<br />
R E M O T E S T A R T<br />
Monitor Mode<br />
By pressing the TEMPERATURE-POWER-SPEED push-button, the<br />
display shows:<br />
P = X X X W T = X X ° C<br />
I = X : X A X X X X H z<br />
where<br />
• P = is the power supplied by the controller to the pump (0 to 999<br />
watt)<br />
• T = is the temperature of the outer ring of the upper bearing<br />
(range 00 to 99 °C)<br />
• I = current<br />
• HHH = Indicates the measurement unit set for the threshold (Hz,<br />
mA, sec. W)<br />
• XXXXHz = is the excitation frequency in hertz<br />
1 Press another time to exit from this menu.<br />
2 Press the CYCLE NUMBER once and the display shows:<br />
X X X X C Y C L E X X X X X m<br />
P U M P L I F E X X X X X h<br />
Where:<br />
• XXXXCYCLE = are the numbers of cycles performed (range 0 to<br />
9999)<br />
• XXXXXm = is the elapsed time related to the last cycle (range 0<br />
to 99999 minutes)<br />
• PUMP LIFE XXXXXh = is the total operation time of the pump<br />
(range 0 to 99999 hours).<br />
118/138 PALM-I UNIT User Manual / 87-900-950-01 (E)