18<br />

Technical Information<br />

Use<br />

and finally the pump status (for example):<br />

P U M P R E A D Y : U S E<br />

R E M O T E S T A R T<br />

Monitor Mode<br />

By pressing the TEMPERATURE-POWER-SPEED push-button, the<br />

display shows:<br />

P = X X X W T = X X ° C<br />

I = X : X A X X X X H z<br />

where<br />

• P = is the power supplied by the controller to the pump (0 to 999<br />

watt)<br />

• T = is the temperature of the outer ring of the upper bearing<br />

(range 00 to 99 °C)<br />

• I = current<br />

• HHH = Indicates the measurement unit set for the threshold (Hz,<br />

mA, sec. W)<br />

• XXXXHz = is the excitation frequency in hertz<br />

1 Press another time to exit from this menu.<br />

2 Press the CYCLE NUMBER once and the display shows:<br />

X X X X C Y C L E X X X X X m<br />

P U M P L I F E X X X X X h<br />

Where:<br />

• XXXXCYCLE = are the numbers of cycles performed (range 0 to<br />

9999)<br />

• XXXXXm = is the elapsed time related to the last cycle (range 0<br />

to 99999 minutes)<br />

• PUMP LIFE XXXXXh = is the total operation time of the pump<br />

(range 0 to 99999 hours).<br />

118/138 PALM-I UNIT User Manual / 87-900-950-01 (E)