18<br />
Technical Information<br />
Error Messages<br />
Error Messages<br />
For a certain type of failure, the controller wills self-diagnose the<br />
error and the following messages will be displayed.<br />
NOTE<br />
If the pump is not connected, the display will be as shown in the following<br />
figure.<br />
C H E C K C O N N E C T I<br />
T O P U M P<br />
Check connection between controller and pump, then press START<br />
RESET pushbutton twice to start the pump.<br />
O<br />
N<br />
NOTE<br />
If the external interlock connections are open, when the START pushbutton is<br />
pressed the display will be as shown in the following figure.<br />
P U M P W A I T I N G<br />
I N T E R L O C K<br />
Provide a right interlock connection.<br />
NOTE<br />
If the upper bearing/pump temperature exceed the maximum permissible, the<br />
pump is shut off, and the display will be as shown in the following figure.<br />
F A U L T : P U M P<br />
O V E R T E M P .<br />
The message will stay on until the temperature decreases below<br />
threshold value. Press the STOP RESET pushbutton twice to start the<br />
pump.<br />
NOTE<br />
If the controller temperature exceeds its limit, the pump is shut off, and the<br />
display will be as shown in the following figure.<br />
130/138 PALM-I UNIT User Manual / 87-900-950-01 (E)