Technical Information<br />
Use<br />
18<br />
N O R M A L O P E R A T I O N<br />
During acceleration of the pump or during any operating condition, it<br />
is always possible to select the other parameters to be displayed<br />
(TEMPERATURE POWER or CYCLE NUMBER push buttons).<br />
If this is the case when the pump reaches the normal speed, the<br />
display reverts to the previous figure.<br />
Self Diagnostic Mode<br />
The Self-Diagnostic Mode is able to test the controller efficiency.<br />
NOTE<br />
The self-diagnostic mode isn’t only a PALM-I feature; a controller’s SW able to<br />
provide diagnostic functions is needed.<br />
To enter in this mode press the CYCLE NUMBER… and the<br />
CONTROLLER INFORMATION buttons together for few seconds.<br />
NOTE<br />
The self-diagnostic mode is operative only if the controller is in STOP or FAIL<br />
status. When in the START mode, the pump will be stopped.<br />
The display shows:<br />
A U T O M A T I C T E S T ?<br />
X X X<br />
Where:<br />
XXX set automatic test (YES/NO) by CYCLE NUMBER button then<br />
press TEMPERATURE button to execute all the possible tests (YES)<br />
or skip the automatic test (NO). During the automatic test the PALM-I<br />
shows the single function under test and gives the result: PASS or<br />
FAIL. If PASS automatically tests the next available function, if FAIL<br />
it stops; press TEMPERATURE button to continue.<br />
If the PALM-I display<br />
PALM-I UNIT User Manual / 87-900-950-01 (E) 125/138