Technical Information<br />
Use<br />
18<br />
Pressing the TEMPERATURE push button will display:<br />
P U M P L I F E R E S E T<br />
S E L E C T I O N : X X X<br />
where:<br />
• Selection: XXX reset the pump lifetime (yes/no).<br />
Pressing the TEMPERATURE push button will display:<br />
I N T E R L O C K T Y P E<br />
S E L E C T I O N : X X X X<br />
where:<br />
• XXXX select the type of interlock check CONT./IMP.<br />
(continuos/impulsive).<br />
Pressing the TEMPERATURE push button will display:<br />
H I G H S P E E D S E T<br />
where:<br />
X X X X H z<br />
• XXXX set the high-speed excitation frequency in hertz.<br />
Pressing the TEMPERATURE push button will display:<br />
M A X H I G H S P E E D<br />
where:<br />
X X X X H z<br />
• Selection: X selects the maximum permissible value of highspeed<br />
excitation frequency in hertz.<br />
Pressing the TEMPERATURE push button will display:<br />
V E N T V A L V E<br />
S E T T I N G : X X X<br />
where:<br />
PALM-I UNIT User Manual / 87-900-950-01 (E) 123/138