18<br />
Technical Information<br />
Use<br />
A U T O D I A G N O T<br />
A V A I L A B L E<br />
This means the controller’s SW don’t provides self-diagnostic<br />
function and an upgrade is needed.<br />
If the automatic test is skipped, the display shows:<br />
R S 2 3 2 P O R T<br />
NOTE<br />
Provided below are all the tests that can be performed. Some of these are only<br />
available for specific controller models.<br />
This proposes a serial RS-232 test; pressing the CYCLE NUMBER<br />
button the test starts and the result is given (PASS or FAIL). To skip<br />
this test press TEMPERATURE button and display shows:<br />
R S 4 8 5 P O R T<br />
This proposes a serial RS-485 test; pressing the CYCLE NUMBER…<br />
button the test starts and the result is given (PASS OR FAIL). To skip<br />
this test press TEMPERATURE button.<br />
NOTE<br />
The PALM-I tries to connect the controller by RS232 first if this don’t work it<br />
tries with RS485 but don’t show any error message.<br />
The display shows:<br />
S E T P O I N T O U T<br />
This proposes setpoint out test; pressing the CYCLE NUMBER<br />
button the test starts and the result is given (PASS OR FAIL). To skip<br />
this test press TEMPERATURE button.<br />
The display shows:<br />
126/138 PALM-I UNIT User Manual / 87-900-950-01 (E)