2.2 Response formats<br />

All <strong>responses</strong> sent by the Halogen Moisture Analyzer to the transmitter (e.g. computer) to<br />

acknowledge the received comm<strong>and</strong> have one of the following formats:<br />

• Response with weight value<br />

• Response without weight value<br />

• Error message<br />

2.2.1 Format of the response with weight value<br />

A general description of the response with weight value is the following.<br />

ID / Status / WeightValue / Unit C R L F<br />

1 character 11 characters 1 – X characters<br />

1 – 2 characters<br />

ID<br />

Response identification<br />

/ Space (ASCII 32 dec.)<br />

Status<br />

WeightValue<br />

Unit<br />

C R<br />

L F<br />

Interface status of the Halogen Moisture Analyzer, see description of the<br />

comm<strong>and</strong>s <strong>and</strong> <strong>responses</strong><br />

Weighing result; shown as number with 11 digits, incl. decimal point<br />

<strong>and</strong> sign – directly in front of the first digit if value negative. The weight<br />

value appears right-aligned. Preceding zeros are not shown with the<br />

exception of the zero to the left of the decimal point.<br />

Weight unit<br />

Carriage Return (ASCII 13 dec.)<br />

Line Feed (ASCII 10 dec.)<br />

Comment<br />

C R<br />

L F<br />

will not be shown in this description.<br />

Example<br />

Response with stable weight value of 0.256 g:<br />

S/S///////0.256/g<br />