3.1 Commands and responses MT-SICS level 0 - Mettler Toledo
3.1 Commands and responses MT-SICS level 0 - Mettler Toledo
3.1 Commands and responses MT-SICS level 0 - Mettler Toledo
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
HA23<br />
Comm<strong>and</strong><br />
Response<br />
Inquiry of last temperature adjustment<br />
HA23 Inquiry of the last successful temperature adjustments<br />
HA23/A/x1/x2/x3/x4/x5/x6<br />
x1 Number of successful adjustments<br />
x2 Day of the last successful adjustment<br />
x3 Month of the last successful adjustment<br />
x4 Year of the last successful adjustment<br />
x5 Hour of the last successful adjustment<br />
x6 Minute of the last successful adjustment<br />
Example<br />
Comm<strong>and</strong><br />
Response<br />
HA23<br />
HA23/A/15/29/02/1996/09/34<br />
A total of 15 successful temperature adjustments have been<br />
performed. The last took place on February 29, 1996 at 9.34.<br />
Comments<br />
• The time of the last successful temperature adjustment is specified in the 24-hour format.<br />
• Possible years are 1970 … 2037.<br />
• The counter for the temperature adjustments runs to 65535.<br />
HA24 – Inquiry of temperature<br />
Comm<strong>and</strong> HA24 Inquiry of current temperature<br />
Response HA24/A/x Current temperature in °C<br />
Example<br />
Comm<strong>and</strong><br />
Response<br />
HA24 Inquiry of current temperature<br />
HA24/A/105 The temperature is 105 °C.<br />
34