21.11.2014 Views

3.1 Commands and responses MT-SICS level 0 - Mettler Toledo

3.1 Commands and responses MT-SICS level 0 - Mettler Toledo

3.1 Commands and responses MT-SICS level 0 - Mettler Toledo

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

SIR<br />

Send weight value immediately <strong>and</strong> repeat<br />

Comm<strong>and</strong><br />

Response<br />

SIR Send the weight values repeatedly, irrespective of balance stability<br />

S/S/WeightValue/Unit<br />

Stable weight value<br />

S/D/WeightValue/Unit<br />

Nonstable (dynamic) weight value<br />

Example<br />

Comm<strong>and</strong><br />

SIR<br />

Send current weight values at intervals<br />

Response<br />

S/D///////2.907/g<br />

S/D///////2.850/g<br />

S/S///////2.797/g<br />

S/S///////2.775/g<br />

S/D///////2.770/g<br />

…<br />

The Halogen Moisture Analyzer sends stable or nonstable weight<br />

values at intervals of 150 ms<br />

Comment<br />

For details see Reference manual 705184.<br />

Z<br />

Zero<br />

Comm<strong>and</strong><br />

Response<br />

Z<br />

Z/A<br />

Zero the Halogen Moisture Analyzer<br />

Zero setting performed, i.e. stability criterion <strong>and</strong> zero setting<br />

range complied with<br />

Example<br />

Comm<strong>and</strong><br />

Response<br />

Z<br />

Z/A<br />

Zero<br />

Zero setting performed<br />

Comments<br />

• This comm<strong>and</strong> is equivalent to pressing the Q key.<br />

• For details see Reference manual 705184.<br />

16