3.1 Commands and responses MT-SICS level 0 - Mettler Toledo
3.1 Commands and responses MT-SICS level 0 - Mettler Toledo
3.1 Commands and responses MT-SICS level 0 - Mettler Toledo
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
SIR<br />
Send weight value immediately <strong>and</strong> repeat<br />
Comm<strong>and</strong><br />
Response<br />
SIR Send the weight values repeatedly, irrespective of balance stability<br />
S/S/WeightValue/Unit<br />
Stable weight value<br />
S/D/WeightValue/Unit<br />
Nonstable (dynamic) weight value<br />
Example<br />
Comm<strong>and</strong><br />
SIR<br />
Send current weight values at intervals<br />
Response<br />
S/D///////2.907/g<br />
S/D///////2.850/g<br />
S/S///////2.797/g<br />
S/S///////2.775/g<br />
S/D///////2.770/g<br />
…<br />
The Halogen Moisture Analyzer sends stable or nonstable weight<br />
values at intervals of 150 ms<br />
Comment<br />
For details see Reference manual 705184.<br />
Z<br />
Zero<br />
Comm<strong>and</strong><br />
Response<br />
Z<br />
Z/A<br />
Zero the Halogen Moisture Analyzer<br />
Zero setting performed, i.e. stability criterion <strong>and</strong> zero setting<br />
range complied with<br />
Example<br />
Comm<strong>and</strong><br />
Response<br />
Z<br />
Z/A<br />
Zero<br />
Zero setting performed<br />
Comments<br />
• This comm<strong>and</strong> is equivalent to pressing the Q key.<br />
• For details see Reference manual 705184.<br />
16