HA01<br />

Reset application / escape<br />

Comm<strong>and</strong><br />

Response<br />

HA01<br />

HA01/A<br />

Reset application / escape<br />

Application reset<br />

Comment<br />

This comm<strong>and</strong> has the same effect as the Reset key, see operating instructions of the Halogen<br />

Moisture Analyzer. It terminates all current comm<strong>and</strong>s <strong>and</strong> activities.<br />

HA02<br />

Set factory settings<br />

Comm<strong>and</strong><br />

Response<br />

HA02<br />

HA02/A<br />

Set factory setting of the menu <strong>and</strong> method parameters<br />

Mean <strong>and</strong> method parameters set to factory setting<br />

Comment<br />

This comm<strong>and</strong> terminates a drying.<br />

HA03<br />

Switch keypad on/off<br />

Comm<strong>and</strong> HA03/x x = 0 Keypad of Halogen Moisture Analyzer switched off<br />

x = 1 Keypad of Halogen Moisture Analyzer switched on (factory<br />

setting)<br />

Response HA03/A<br />

HA03/L<br />

Comm<strong>and</strong> executed<br />

Comm<strong>and</strong> understood, parameter wrong<br />

Comment<br />

This comm<strong>and</strong> has no effect on the on/off key. The user can thus switch the instrument on or<br />