3.1 Commands and responses MT-SICS level 0 - Mettler Toledo
3.1 Commands and responses MT-SICS level 0 - Mettler Toledo
3.1 Commands and responses MT-SICS level 0 - Mettler Toledo
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
HA01<br />
Reset application / escape<br />
Comm<strong>and</strong><br />
Response<br />
HA01<br />
HA01/A<br />
Reset application / escape<br />
Application reset<br />
Comment<br />
This comm<strong>and</strong> has the same effect as the Reset key, see operating instructions of the Halogen<br />
Moisture Analyzer. It terminates all current comm<strong>and</strong>s <strong>and</strong> activities.<br />
HA02<br />
Set factory settings<br />
Comm<strong>and</strong><br />
Response<br />
HA02<br />
HA02/A<br />
Set factory setting of the menu <strong>and</strong> method parameters<br />
Mean <strong>and</strong> method parameters set to factory setting<br />
Comment<br />
This comm<strong>and</strong> terminates a drying.<br />
HA03<br />
Switch keypad on/off<br />
Comm<strong>and</strong> HA03/x x = 0 Keypad of Halogen Moisture Analyzer switched off<br />
x = 1 Keypad of Halogen Moisture Analyzer switched on (factory<br />
setting)<br />
Response HA03/A<br />
HA03/L<br />
Comm<strong>and</strong> executed<br />
Comm<strong>and</strong> understood, parameter wrong<br />
Comment<br />
This comm<strong>and</strong> has no effect on the on/off key. The user can thus switch the instrument on or<br />
off even if the keypad is switched off.<br />
28