HA27<br />

Inquiry of drying result<br />

Comm<strong>and</strong><br />

HA27/x1<br />

Inquiry of drying data in configurable display mode<br />

x1 = 0 currently set display mode<br />

x1 = 1 Grams<br />

x1 = 2 DC (dry content)<br />

x1 = 3 MC (moisture content), (factory setting)<br />

x1 = 4 AM (ATRO moisture content)<br />

x1 = 5 AD (ATRO dry content)<br />

Response<br />

HA27/A/x1/x2<br />

x1 Drying status (always 7 digit number)<br />

x2 Display mode (g, %DC, %MC, %AM, %AD)<br />

HA27/I Response not available (drying in progress)<br />

Example<br />

Comm<strong>and</strong><br />

Response<br />

HA27/3 Inquiry of drying result<br />

HA27/A//<br />

//-73.25%MC<br />

Drying result -73.25 % MC<br />

Comment<br />

If the measuring results exceed the tolerances for ATRO result display (L-999.99 % AM or<br />

> 999.99 % AD) the selected results in x1 = 4 AM or x1 = 5 AD will automatically be<br />

transferred in x1 = 3 MC or x1 = 2 DC respectively.<br />