Parameters<br />

x14 Time unit for free switch-off criterion<br />

Possible values: 5 … 180 s, factory setting: 50 s<br />

x15 Free print interval<br />

x15 = 0 No free print interval, only the preset print intervals<br />

are available (factory setting)<br />

x15 = 1 User-defined print interval<br />

x16 Entering free print interval<br />

Possible values: 5 … 3600 s, factory setting: 900 s<br />

x17 Working with methods<br />

x17 = 0 Working without methods<br />

(factory setting with HG53)<br />

x17 = 1 Working with methods<br />

(factory setting with HR73)<br />

This setting is not possible with the HG53<br />

Example<br />

Comm<strong>and</strong><br />

Response<br />

HA41/1/0/2/0/1/0/10/1/1/"<strong>Mettler</strong>-<strong>Toledo</strong> GmbH"<br />

/1/1/10/50/1/900/1<br />

The following menu parameters are set differently from the<br />

factory setting:<br />

• Record of company name switched on<br />

• Company name: <strong>Mettler</strong>-<strong>Toledo</strong> GmbH<br />

• Free switch-off criterion switched on with factory setting for<br />

weight loss <strong>and</strong> time unit<br />

• Free print interval switched on with factory setting for the free<br />

print interval<br />

HA41/A<br />

Menu parameters set<br />

Comment<br />

Setting the menu parameters terminates a drying.<br />