Parameters<br />
x14 Time unit for free switch-off criterion<br />
Possible values: 5 … 180 s, factory setting: 50 s<br />
x15 Free print interval<br />
x15 = 0 No free print interval, only the preset print intervals<br />
are available (factory setting)<br />
x15 = 1 User-defined print interval<br />
x16 Entering free print interval<br />
Possible values: 5 … 3600 s, factory setting: 900 s<br />
x17 Working with methods<br />
x17 = 0 Working without methods<br />
(factory setting with HG53)<br />
x17 = 1 Working with methods<br />
(factory setting with HR73)<br />
This setting is not possible with the HG53<br />
Example<br />
Comm<strong>and</strong><br />
Response<br />
HA41/1/0/2/0/1/0/10/1/1/"<strong>Mettler</strong>-<strong>Toledo</strong> GmbH"<br />
/1/1/10/50/1/900/1<br />
The following menu parameters are set differently from the<br />
factory setting:<br />
• Record of company name switched on<br />
• Company name: <strong>Mettler</strong>-<strong>Toledo</strong> GmbH<br />
• Free switch-off criterion switched on with factory setting for<br />
weight loss <strong>and</strong> time unit<br />
• Free print interval switched on with factory setting for the free<br />
print interval<br />
HA41/A<br />
Menu parameters set<br />
Comment<br />
Setting the menu parameters terminates a drying.<br />