D<br />
Display<br />
Write into display<br />
Comm<strong>and</strong> D/"TEXT"<br />
Response D/A<br />
D/R<br />
D/I<br />
D/L<br />
Write TEXT into Halogen Moisture Analyzer display<br />
TEXT appears unabridged left-aligned in the Halogen Moisture<br />
Analyzer display marked by the symbol *<br />
The end of the text appears in the Halogen Moisture Analyzer<br />
display, the start is cut off <strong>and</strong> text is marked by the symbol *<br />
Comm<strong>and</strong> not executable<br />
Comm<strong>and</strong> understood, parameter wrong<br />
Example<br />
Comm<strong>and</strong><br />
Response<br />
D/"HALLO"<br />
D/A<br />
Write HALLO into the Halogen Moisture Analyzer display<br />
The full text HALLO appears in the Halogen Moisture Analyzer<br />
display<br />
Clear display<br />
Comm<strong>and</strong> D/""<br />
Response D/A<br />
Clear Halogen Moisture Analyzer display<br />
Display cleared, marked by the symbol *<br />
Comments<br />
• Max. 20 characters are admissible for "text".<br />
• The following ASCII character set applies to "text":<br />
0…9, A…Z, a…z, #, &, *, +, -, ., /, [, ], /, :.<br />
• A display comm<strong>and</strong> can be cleared with the Reset key.<br />
• This comm<strong>and</strong> can be executed only in the instrument statuses "basic mode", "ready for<br />
taring", "weighing-in" <strong>and</strong> "ready for start".<br />
• For details see Reference manual 705184.<br />
DW – Weight display (Display show Weight)<br />
Comm<strong>and</strong><br />
Response<br />
DW<br />
DW/A<br />
DW/I<br />
Switch display to weight mode<br />
Display shows the current weight value<br />
Comm<strong>and</strong> not executable<br />
Comment<br />
This comm<strong>and</strong> can be executed only in the instrument statuses "basic mode", "ready for<br />
taring", "weighing-in" <strong>and</strong> "ready for start".<br />