LC/MS <strong>System</strong> Configuration <strong>Test</strong>— AB SCIEX TripleTOF 5600 <strong>System</strong><br />
• In the table, specify the m/z of this peptide along with the collision energy (45)<br />
and a CES of 3 V.<br />
• The Declustering Potential is 70.<br />
11. Click Finish to save the method.<br />
Add LC Information to the Acquisition Method<br />
If the autosampler and pump method information was not specified above, add it now.<br />
Note: The Q1 Isolation window is automatically set to Unit. This is the correct setting for<br />
this test.<br />
1. Click Acquisition Method in the left pane, and then select LC Sync as the<br />
Synchronization Mode.<br />
2. Click <strong>Eksigent</strong> AS2 and then select the autosampler method, AS2 10uLloop 1uLinj<br />
nanoflex trap elute.ini or AS2 10uLloop 1uLinj trap elute.ini, depending on the<br />
installation.<br />
Figure 1-11 Software Application Properties tab—autosampler filename<br />
3. Click <strong>Eksigent</strong> Gradient 1, and then select the gradient 2 pump method, CH2 15min<br />
400nLmin nanoflex trap.ini or CH2 15min 400nLmin trap.ini, depending on the<br />
installation.<br />
4. Click <strong>Eksigent</strong> Loading Pump, and then select the loading pump method, Load<br />
Pump 2 min Trap Wash.ini.<br />
5. Save the method as “<strong>System</strong> Functional <strong>Test</strong>”.<br />
<strong>NanoLC</strong> Ultra ® <strong>System</strong><br />
24 of 124<br />
<strong>System</strong> <strong>Integration</strong> <strong>Test</strong><br />
D5030175 A