LC/MS <strong>System</strong> Configuration <strong>Test</strong>— AB SCIEX TripleTOF 5600 <strong>System</strong><br />

Figure 1-14 LC Method Settings dialog—Run Conditions tab<br />

6. Click Save.<br />

Create the Gradient Method in the <strong>Eksigent</strong> Control Software<br />

For the analytical gradient (typically on the Gradient 2 pump with the nanoflow module), create<br />

the gradient method.<br />

1. Make sure that Gradient 2 is selected as the channel (top, right corner of window).<br />

2. Click LC Methods.<br />

3. If this is a Nanoflex system installation, then in the Name field, type CH2 45min<br />

300nL/min Nanoflex trap, and click Save.<br />

Or<br />

If this is a column installation, then in the Name field, type CH2 45min 300nLmin<br />

column trap, and click Save<br />

4. On the Gradient Table tab, specify the method, as shown in Figure 1-15.<br />

<strong>NanoLC</strong> Ultra ® <strong>System</strong><br />

28 of 124<br />

<strong>System</strong> <strong>Integration</strong> <strong>Test</strong><br />

D5030175 A