LC/MS <strong>System</strong> Configuration <strong>Test</strong>— AB SCIEX TripleTOF 5600 <strong>System</strong><br />

Figure 1-26 TOF MS XICs of 5 peptides from Beta-Galactosidase <strong>NanoLC</strong> Ultra trap elute<br />

run (200 fmol on column)<br />

Note: This test is not a specification. Use this example to confirm injection<br />

and peak separation and shape only.<br />

6. Make sure the peaks have good separation and shape.<br />

i. In the PeakView software, on the Window menu, click Graph Selection<br />

Window.<br />

ii. To collect the Peak Area for each XIC, click each (color-keyed) line in the<br />

upper left.<br />

The retention time appears on the top of the peak selected. The bottom line of<br />

the Graph Selection Info displays the Peak Area.<br />

Measure the Performance of the TOF MS/MS Scan<br />

1. In the main TIC pane, extract individual TICs by double-clicking the magenta arrow.<br />

2. Select the MS/MS scan, and then select Remove all Traces Except Active to<br />

display the pane shown at the bottom of Figure 1-27.<br />

<strong>NanoLC</strong> Ultra ® <strong>System</strong><br />

40 of 124<br />

<strong>System</strong> <strong>Integration</strong> <strong>Test</strong><br />

D5030175 A