LC/MS <strong>System</strong> Configuration <strong>Test</strong>— AB SCIEX TripleTOF 5600 <strong>System</strong><br />

<strong>Test</strong> Results—TripleTOF 5600 Instruments<br />

For TripleTOF 5600 instruments, complete this table with the results from the ion from the Beta-<br />

Galactosidase digest solution stock. Make sure that the peak area is within specification.<br />

Most of the XICs obtained should have peak widths of ~0.17 minute half height on average and<br />

have peak intensities similar to that shown in Figure 1-29. Some peaks will be narrower and<br />

some will be broader Additionally, verify that the peaks elute within 5 minutes of each other.<br />

Beta-Gal Lot Number: ___________________________________<br />

Table 1-4 LC/MS Specification <strong>Test</strong>—TripleTOF 5600 Instruments<br />

Q1 Peptide ID Spec.<br />

(peak area)<br />

503.2368 YSQQQLMETSHR 1.0E+05<br />

542.2645 GDFQFNISR 1.0E+05<br />

671.3379 VDEDQPFPAVPK 2.0E+05<br />

714.8469 DWENPGVTQLNR 5.0E+04<br />

729.3652 APLDNDIGVSEATR 2.0E+05<br />

Specification Passed<br />

Notes<br />

Actual<br />

<br />

<strong>NanoLC</strong> Ultra ® <strong>System</strong><br />

50 of 124<br />

<strong>System</strong> <strong>Integration</strong> <strong>Test</strong><br />

D5030175 A