LC/MS <strong>System</strong> Configuration <strong>Test</strong>— AB SCIEX TripleTOF 5600 <strong>System</strong><br />
Figure 1-24 Add Samples dialog—<strong>System</strong> Functional <strong>Test</strong><br />
3. Click OK.<br />
4. Save the data file as TT5600 system LC BGal functional status check .<br />
5. On the Location tab, specify the location of the Beta-Galactosidase sample in the<br />
autosampler.<br />
Run the Batch<br />
1. On the Submit tab, click Submit.<br />
2. In the View menu, click Sample Queue.<br />
3. In the Acquire menu, click Start Sample.<br />
Verify the Results in the PeakView Software<br />
1. Open the data file and double-click the magenta arrow to display the individual TICs.<br />
2. Right-click the TIC of the TOF MS and select Remove all Traces Except Active to<br />
display the overlaid TOF MS TIC.<br />
3. Extract the TOF MS peak XICs for the target peptides.<br />
i. On the Show menu, click Extract Ions Using.<br />
ii. Specify the masses and extraction width for the peptides as shown in<br />
Figure 1-25. Only the masses and widths are required.<br />
<strong>NanoLC</strong> Ultra ® <strong>System</strong><br />
38 of 124<br />
<strong>System</strong> <strong>Integration</strong> <strong>Test</strong><br />
D5030175 A