4.0 Operation<br />
4.5 Ear rinsing module 37 °C basic<br />
Wh<strong>en</strong> switching on <strong>the</strong> unit, <strong>the</strong> quivers for <strong>the</strong> ear rinsing system will be<br />
activated and heated up to 37°C.<br />
Only use antibacterial rinsing fluid for rinsing <strong>the</strong> auditory channel.<br />
The temperature of <strong>the</strong> rinsing fl uid has to be checked with a <strong>the</strong>rmometer before using.<br />
It must be 37°C ± 2°C, o<strong>the</strong>rw<strong>is</strong>e <strong>the</strong> pati<strong>en</strong>t may feel dizzy or can collapse.<br />
The vacuum gauge may indicate max. -0.4 bar at closed op<strong>en</strong>ing on . It has to be checked before each<br />
application on <strong>the</strong> pati<strong>en</strong>t.<br />
To avoid awkward and loud side-air-tones, <strong>the</strong> application part `rinsing peak` has to be held close to <strong>the</strong><br />
acoustic meatus.<br />
Rinsing fl uid, which <strong>is</strong> no longer in use, has to be d<strong>is</strong>posed at <strong>the</strong> <strong>en</strong>d of <strong>the</strong> practice business.<br />
<br />
<br />
<br />
<br />
<br />
<br />
• The rinsing bottle will be fi lled with warm rinsing fl uid<br />
(37°C ± 2°C). The rinsing lid with <strong>the</strong> hose will be<br />
screwed onto <strong>the</strong> bottle and slightly tight<strong>en</strong>ed.<br />
• Via <strong>the</strong>rmometer <strong>the</strong> temperature can be monitored.<br />
• The t<strong>en</strong>der peak will be attached onto <strong>the</strong> cylindric<br />
attempt at <strong>the</strong> rinsing lid as far as it will go<br />
• Connect <strong>the</strong> suction hose with <strong>the</strong> vacuum port of <strong>the</strong><br />
ear rinsing system. Set suction system on maximum<br />
vacuum.<br />
• Now <strong>the</strong> ear rinsing system with <strong>the</strong> t<strong>en</strong>der peak <strong>is</strong><br />
putted into <strong>the</strong> acoustic meatus. Keep <strong>the</strong> side-air op<strong>en</strong>ing<br />
closed with your fi nger. As long as <strong>the</strong> side-air op<strong>en</strong>ing<br />
<strong>is</strong> op<strong>en</strong>, it will be rinsed.<br />
Ear rinsing system<br />
t<strong>en</strong>der peak for rinsing<br />
side-air aperture<br />
vacuum port on system<br />
rinsing lid (complete)<br />
rinsing bottle<br />
safety aperture<br />