3.0 Assembly: Connecting conditions<br />
3.3 Water connection<br />
!<br />
Requirem<strong>en</strong>ts for water supply:<br />
• Easily accessible water tap with hose connection G 3/4“.<br />
• Connection including pipe aeration <strong>is</strong> recomm<strong>en</strong>ded.<br />
• The water which <strong>is</strong> provided by <strong>the</strong> household connection must at least meet <strong>the</strong> WHO guidelines or <strong>the</strong><br />
country-specific guidelines for drinking water.<br />
• Rinse water supply line in order to remove any contamination from <strong>the</strong> system.<br />
• Wh<strong>en</strong> clean water <strong>is</strong> available, connect delivery hose to water tap m<strong>en</strong>tioned above.<br />
• Required pressure in domestic water system: >2 bar, but 3.8 >21.3<br />
• The use of demineral<strong>is</strong>ed water may affect <strong>the</strong> function of <strong>the</strong> waste water d<strong>is</strong>posal!<br />
3.4 Water drain<br />
Requirem<strong>en</strong>ts:<br />
• Perman<strong>en</strong>tly installed connection fitting with G3/4“ external thread.<br />
An anti-syphon trap <strong>is</strong> integrated in <strong>the</strong> unit!<br />
• Connect waste water hose with unit and <strong>the</strong> G3/4“ connection fitting. Thereby fit <strong>en</strong>closed gaskets into <strong>the</strong><br />
screwing.<br />
1730<br />
740<br />
250<br />
445<br />
640<br />
1550<br />
630<br />
730<br />
19<strong>61</strong><br />
2940<br />
979,5<br />
9