2.0 For your safety<br />
• Only switch on unit at available water consumption, wh<strong>en</strong><br />
water supply <strong>is</strong> guaranteed!<br />
!<br />
For your safety<br />
• The <strong>ATMOS</strong> S <strong>61</strong> <strong>Servant</strong> <strong>ENT</strong> <strong>workstation</strong> unit <strong>is</strong><br />
produced according to IEC 601/ EN 60601 and l<strong>is</strong>ted in<br />
<strong>the</strong> following classes:<br />
• VDE Class of protection 1<br />
• Class IIa (EEC 93/42).<br />
• Correct configuration in assembly of country-specific<br />
connections:<br />
gre<strong>en</strong>/yellow: protective conductor (PE)<br />
blue:<br />
neutral conductor (N)<br />
black resp. brown: phase (L)<br />
• Att<strong>en</strong>tion! Mirrorheaters may g<strong>en</strong>erate temperatures<br />
above 40°C!<br />
• The <strong>ATMOS</strong> S <strong>61</strong> <strong>Servant</strong> <strong>ENT</strong> <strong>workstation</strong> may be<br />
used in superv<strong>is</strong>ed operation by qualified personnel<br />
only which has be<strong>en</strong> author<strong>is</strong>ed by <strong>ATMOS</strong> and which<br />
has be<strong>en</strong> trained for operating <strong>the</strong> appliance (IEC 601-<br />
1/EN 60601-1).<br />
• The mains voltage specified on <strong>the</strong> type plate must<br />
correspond with <strong>the</strong> data of <strong>the</strong> power supply system.<br />
• Make sure prior to every application of <strong>the</strong> equipm<strong>en</strong>t that<br />
it <strong>is</strong> technically safe and in proper condition. Damaged<br />
leads and hoses must be replaced immediately!<br />
• D<strong>is</strong>play instrum<strong>en</strong>ts and valves must be checked for<br />
correct function in regular intervals !<br />
• Inspection of compressed air and vacuum d<strong>is</strong>play by<br />
service technician every 2 years !<br />
• Arrange a sight check every morning for <strong>the</strong> hoses,<br />
collection jars and cables.<br />
Exchange damaged parts!<br />
Only use correct and undamaged line voltages<br />
and ext<strong>en</strong>sion cables.<br />
• Switch off <strong>the</strong> unit before op<strong>en</strong> it and separate it<br />
from <strong>the</strong> line voltage.<br />
• Maximum load for clipboards: 2 kg<br />
• Never leave <strong>the</strong> pati<strong>en</strong>t unatt<strong>en</strong>ded at <strong>the</strong> treatm<strong>en</strong>t unit.<br />
• <strong>Th<strong>is</strong></strong> product <strong>is</strong> not re-steril<strong>is</strong>able. Repeated reuse of<br />
compon<strong>en</strong>ts which are marked with a 2 <strong>is</strong> forbidd<strong>en</strong>.<br />
In case of repeated reuse <strong>the</strong>se compon<strong>en</strong>ts lose <strong>the</strong>ir<br />
function and <strong>the</strong>re <strong>is</strong> a high infection r<strong>is</strong>k.<br />
• In case that <strong>the</strong>re <strong>is</strong> no adapter attached to <strong>the</strong> light<br />
conductor, touching <strong>the</strong> light source with ei<strong>the</strong>r <strong>the</strong><br />
fingers or any tools should be avoided. On <strong>the</strong> one hand<br />
th<strong>is</strong> may damage <strong>the</strong> l<strong>en</strong>ses and on <strong>the</strong> o<strong>the</strong>r hand <strong>the</strong>re<br />
<strong>is</strong> a r<strong>is</strong>k of injury.<br />
• The <strong>ENT</strong> unit requires clean water (drinking-water quality)<br />
for <strong>the</strong> operation. In case <strong>the</strong> clean water cannot be provided<br />
by <strong>the</strong> water supply, a pre-filter has to be installed.<br />
The relevant country specific regulations for <strong>the</strong> instal-lation<br />
have to be considered !<br />
• Please note:<br />
A medical insulating transformer with earth leakage<br />
monitor or any similar safety system acc. to EN 60 601-1<br />
<strong>is</strong> required, if several devices are connected over one<br />
common power supply. The transformer must<br />
correspond to <strong>the</strong> power consumption of all <strong>the</strong> devices<br />
to be connected.<br />
• Do not place used contaminated instrum<strong>en</strong>ts on <strong>the</strong><br />
<strong>ENT</strong> unit except on destined places!<br />
• The ambi<strong>en</strong>t conditions specified in section "Technical<br />
specifications" must be strictly observed!<br />
• Switch off main switch after fin<strong>is</strong>hing work in practice and<br />
close water supply, if pres<strong>en</strong>t.<br />
• Att<strong>en</strong>tion wh<strong>en</strong> working with <strong>en</strong>doscopes near <strong>the</strong> light<br />
sources. Do not look directly into <strong>the</strong> cone of light!<br />
At a conting<strong>en</strong>t light blackout remove <strong>the</strong> <strong>en</strong>doscope out<br />
of <strong>the</strong> area!<br />
• The <strong>ATMOS</strong> S <strong>61</strong> <strong>Servant</strong> may be operated only in<br />
rooms used for medical purposes, but not in areas<br />
subject to explosion hazards and in oxyg<strong>en</strong> rich<br />
<strong>en</strong>vironm<strong>en</strong>ts.<br />
• The <strong>ATMOS</strong> S <strong>61</strong> <strong>Servant</strong> fully complies with <strong>the</strong> electromagnetic<br />
immunity requirem<strong>en</strong>ts of standard IEC 601-1-2<br />
/ EN 60601-1-2 "Electromagnetic compatibility - Medical<br />
Electrical Equipm<strong>en</strong>t".<br />
• The <strong>ATMOS</strong> S <strong>61</strong> <strong>Servant</strong> may not be operated with<br />
units not complying with <strong>the</strong> requirem<strong>en</strong>ts of standard EN<br />
60601-1 "Medical Electrical Equipm<strong>en</strong>t" and EN 60601-<br />
1-2 "Electromagnetic compatibility (Medical Electrical<br />
Equipm<strong>en</strong>t)".<br />
• <strong>ATMOS</strong> <strong>is</strong> not liable for personal injury and damage to<br />
property if<br />
• no original <strong>ATMOS</strong> parts are being used,<br />
• <strong>the</strong> advice for use in <strong>the</strong>se operating instructions <strong>is</strong> not<br />
being observed,<br />
• assembly, new settings, alterations, ext<strong>en</strong>sions and<br />
repairs have be<strong>en</strong> carried out by personnel not author<strong>is</strong>ed<br />
by <strong>ATMOS</strong>.<br />
• These operating instructions correspond with <strong>the</strong> construction<br />
of <strong>the</strong> unit and with <strong>the</strong> curr<strong>en</strong>t status of safetyrelated<br />
standards at <strong>the</strong> time of printing. Proprietary rights<br />
are ex<strong>is</strong>ting for all described circuits, processes, names,<br />
software programs and units..<br />
• Do not constrict <strong>the</strong> air supply on <strong>the</strong> back of <strong>the</strong> unit!<br />
• The ear rinsing modlue 37°C basic must not be actuated<br />
with a 55l-aspiration system!<br />
• Prior to starting <strong>the</strong> treatm<strong>en</strong>t application parts have to be<br />
checked regarding <strong>the</strong>ir temperature.<br />
• It <strong>is</strong> not allowed to use flammable substances with <strong>the</strong><br />
device.<br />
7