3. Commissioning<br />

3.1. Safety Instructions and Principles<br />

Caution in the case of permanent installations! Potential risk of<br />

injury and material damage:Only appropriately trained and qualified<br />

electricians are permitted to connect three-phase equipment to the<br />

power supply.<br />

Please keep to the specifications of the actual version of EN<br />

61010-ff, if you integrate the thermostat (i.e. without plug) into<br />

your electrical installation.<br />

Hard-wired units and Polyphase-Units have to have a switch or a<br />

circuit breaker as separation device.<br />

It is your obligation:<br />

- to install a fuse-protected switch or circuit breaker according to<br />

the data sheet in the installation of your building.<br />

- to place this near to the unit so that it is easily reached by the<br />

user.<br />

- to precisely mark this switch as separation device for this unit.<br />

Units that are connected to the electricity via plug:<br />

The cable must not be longer than 3 m (VDE 0411, part 100).<br />

The connection of the three-phase unit to the electricity or the<br />

change of voltage of the unit must only be carried out by a<br />

competent electrician.<br />

Check that the rating of the fuse protection, the power capacities<br />

and voltages all comply with the requirements on the nameplate<br />

and the technical data sheet.<br />

Caution! Potential risk of injury and material damage:<br />

Please refer to the Data Sheet (appendix) for information on the<br />

safety classification of your thermostat (FL/NFL).<br />

Requirements for thermal fluids classified as NFL:<br />

Only non-flammable thermal fluids are permissible for unattended<br />

operation!<br />

Requirements for thermal fluids classified as FL:<br />

EN 61010-1: Maximum permissible working temperature 25 °C<br />

below the flash point!<br />

Maximum viscosity at the lowest working temperature: 50 mm 2 /s.<br />

Maximum density of the thermal fluid: 1 kg/dm³.<br />

Possible thermoregulation range within the range of the planned<br />

minimum and maximum working temperature.<br />

Do not use thermal fluids with any of the additives ether, ester or<br />

amine. Do not use demineralized water, mineral water, sea water<br />

or CaCl 2 brines!<br />

Compatibility with the materials of the thermostat (stainless steel<br />

1.4301 (V2A) and Viton) and all materials of the connected<br />

application.<br />

Please Note:<br />

All the safety instructions are vital and must be considered during<br />

the operation of the unit in compliance with the operating instructions.<br />