3. Commissioning<br />
3.1. Safety Instructions and Principles<br />
Caution in the case of permanent installations! Potential risk of<br />
injury and material damage:Only appropriately trained and qualified<br />
electricians are permitted to connect three-phase equipment to the<br />
power supply.<br />
Please keep to the specifications of the actual version of EN<br />
61010-ff, if you integrate the thermostat (i.e. without plug) into<br />
your electrical installation.<br />
Hard-wired units and Polyphase-Units have to have a switch or a<br />
circuit breaker as separation device.<br />
It is your obligation:<br />
- to install a fuse-protected switch or circuit breaker according to<br />
the data sheet in the installation of your building.<br />
- to place this near to the unit so that it is easily reached by the<br />
user.<br />
- to precisely mark this switch as separation device for this unit.<br />
Units that are connected to the electricity via plug:<br />
The cable must not be longer than 3 m (VDE 0411, part 100).<br />
The connection of the three-phase unit to the electricity or the<br />
change of voltage of the unit must only be carried out by a<br />
competent electrician.<br />
Check that the rating of the fuse protection, the power capacities<br />
and voltages all comply with the requirements on the nameplate<br />
and the technical data sheet.<br />
Caution! Potential risk of injury and material damage:<br />
Please refer to the Data Sheet (appendix) for information on the<br />
safety classification of your thermostat (FL/NFL).<br />
Requirements for thermal fluids classified as NFL:<br />
Only non-flammable thermal fluids are permissible for unattended<br />
operation!<br />
Requirements for thermal fluids classified as FL:<br />
EN 61010-1: Maximum permissible working temperature 25 °C<br />
below the flash point!<br />
Maximum viscosity at the lowest working temperature: 50 mm 2 /s.<br />
Maximum density of the thermal fluid: 1 kg/dm³.<br />
Possible thermoregulation range within the range of the planned<br />
minimum and maximum working temperature.<br />
Do not use thermal fluids with any of the additives ether, ester or<br />
amine. Do not use demineralized water, mineral water, sea water<br />
or CaCl 2 brines!<br />
Compatibility with the materials of the thermostat (stainless steel<br />
1.4301 (V2A) and Viton) and all materials of the connected<br />
application.<br />
Please Note:<br />
All the safety instructions are vital and must be considered during<br />
the operation of the unit in compliance with the operating instructions.<br />
31-2