Switch on<br />
Open,<br />
Filling<br />
UC016T-3,<br />
UC020T-3,<br />
UC025T-3,<br />
UC040T-3<br />
etc.<br />
also all<br />
UCs with<br />
heating(-H)<br />
Terminate<br />
filling,<br />
close<br />
3. Inbetriebnahme<br />
3.3. Filling Thermofluid<br />
Unichiller (Chiller)<br />
Models: UC016T-3(-H), UC020T-3(-H), UC025T-3(-H),<br />
UC040T-3(-H) etc.<br />
Condition:<br />
make a hose connection to<br />
your external application.<br />
L1 Switch on the main switch.<br />
L2 Lift the bath lid.<br />
L3<br />
L4<br />
Fill in the Thermofluid.<br />
The thermofluid flows from the<br />
expansion vessel into the<br />
thermostat than through the<br />
hose connections to the<br />
external application.<br />
Support the distribution process<br />
with the program „Venting“.<br />
(46-1)<br />
Continue filling procedure and<br />
„Venting“, until Gases are fully<br />
displaced off the System and<br />
the level on the luminescent<br />
display stays constant at 20-<br />
40%.<br />
Optimal level:<br />
20-40%.<br />
L5 The liquid level on the luminescent display stays constant on reaching<br />
of the optimal level.<br />
L6 Stop „Venting“ Program.<br />
L7 Close the bath lid.<br />
Caution! Potencial risk of injury and material damage:<br />
A proper venting is the condition for a troublefree operation!<br />
Commissioning often means a change of liquid.<br />
The manufacturer mostly uses other thermofluids as the user during the<br />
test run. Despite thorough draining, smallest residual could remain in the<br />
circuit and degasify after filling a new fluid and the beginning of the<br />
thermoregulation.<br />
The concequence of the above described and an incomplete venting could<br />
be an increasing presure that leads to sudden discharge of fluid and<br />
vapours!<br />
This danger can be prevented to a large extend with the „Venting“ program<br />
Change of liquid refer to 46-1., program „venting“!<br />
Absolutely pay attention to a thorough degasification at the beginning of<br />
the thermoregulation!<br />