Switch on<br />

Open,<br />

Filling<br />

UC016T-3,<br />

UC020T-3,<br />

UC025T-3,<br />

UC040T-3<br />

etc.<br />

also all<br />

UCs with<br />

heating(-H)<br />

Terminate<br />

filling,<br />

close<br />

3. Inbetriebnahme<br />

3.3. Filling Thermofluid<br />

Unichiller (Chiller)<br />

Models: UC016T-3(-H), UC020T-3(-H), UC025T-3(-H),<br />

UC040T-3(-H) etc.<br />

Condition:<br />

make a hose connection to<br />

your external application.<br />

L1 Switch on the main switch.<br />

L2 Lift the bath lid.<br />

L3<br />

L4<br />

Fill in the Thermofluid.<br />

The thermofluid flows from the<br />

expansion vessel into the<br />

thermostat than through the<br />

hose connections to the<br />

external application.<br />

Support the distribution process<br />

with the program „Venting“.<br />

(46-1)<br />

Continue filling procedure and<br />

„Venting“, until Gases are fully<br />

displaced off the System and<br />

the level on the luminescent<br />

display stays constant at 20-<br />

40%.<br />

Optimal level:<br />

20-40%.<br />

L5 The liquid level on the luminescent display stays constant on reaching<br />

of the optimal level.<br />

L6 Stop „Venting“ Program.<br />

L7 Close the bath lid.<br />

Caution! Potencial risk of injury and material damage:<br />

A proper venting is the condition for a troublefree operation!<br />

Commissioning often means a change of liquid.<br />

The manufacturer mostly uses other thermofluids as the user during the<br />

test run. Despite thorough draining, smallest residual could remain in the<br />

circuit and degasify after filling a new fluid and the beginning of the<br />

thermoregulation.<br />

The concequence of the above described and an incomplete venting could<br />

be an increasing presure that leads to sudden discharge of fluid and<br />

vapours!<br />

This danger can be prevented to a large extend with the „Venting“ program<br />

Change of liquid refer to 46-1., program „venting“!<br />

Absolutely pay attention to a thorough degasification at the beginning of<br />

the thermoregulation!<br />

Please note!<br />

For detailed information on the range of thermofluids incl. technical data<br />

and price quotations, please refer to the Huber catalog or contact your<br />

Huber agent.<br />

33-3<br />

1<br />

L2<br />

L3<br />

L1