Thermal<br />

Fluid<br />

Preparatory<br />

measures<br />

for thermal<br />

fluid<br />

circulation<br />

Control<br />

3. Commissioning<br />

3.2. Preparing the thermostat for use<br />

Applies to bath thermostats<br />

with external pump connection<br />

For internal thermoregulation, please proceed as follows:<br />

close the pump connections with plugs and sleeve nuts (M16x1 / 19<br />

spanner/wrench), at the same time hold against it at the pump<br />

connection with 17 spanner/wrench.<br />

For external thermoregulation, please proceed as follows:<br />

remove the plugs and sleeve nuts on the pump connections.<br />

Replace them with appropriate hose connection to your external<br />

application.<br />

For more information please see chapter 4 „thermoregulation via<br />

controller“.<br />

Applies to chillers (thermostats<br />

with closed external circulation)<br />

2.4Make<br />

sure that the<br />

draining valve is closed.<br />

(upright position or „right<br />

stop“).<br />

2.5 Connect hoses at<br />

connecting pieces 10 and<br />

11 for circulating thermal<br />

fluid to and from the<br />

external device (if fitted).<br />

Note:<br />

Recommended tightening torques for<br />

hose couplings when using silicone<br />

oil as thermal fluid:<br />

M16 x 1 30 Nm<br />

M30 x 1.5 85 Nm<br />

Alternative<br />

Hoses can be connected<br />

via shut-off valves.<br />

(Please refer to your<br />

Huber catalogue under<br />

„accessories“)<br />

Verify the following:<br />

Make sure that all connections are correct and that there are no leaks.<br />

For water-cooled thermostats:<br />

Unobstructed inflow and outflow of the cooling-water exists<br />

Pressure difference between the cooling-water inflow and outflow is 3 to<br />

5 bar. Cooling-water supply temperature is 15 °C to 25 °C.<br />

32-2<br />

12<br />

12<br />

open<br />

close<br />

Hazard!<br />

If fitted, close valves only<br />

when performing work on<br />

the reactor. Remember that<br />

thermal fluid expands and<br />

contracts with changes in<br />

temperature. Sealing the<br />

external application will<br />

expose the application to<br />

these forces!<br />

10 11