Compressor<br />
Auto<br />
Heating<br />
Capacity<br />
Software<br />
version<br />
4. Thermoregulation via<br />
Controller<br />
4.3. 4.2. Editing Wichtige Default Voreinstellungen<br />
Settings<br />
prüfen und ändern<br />
10. Compressor Automatic<br />
11. Maximum 1. Programm Heating 52: Power Rückführung auf Werkseinstellung<br />
12. Software 2. Programm Version90:<br />
Sprachauswahl Digitalanzeige UniCop<br />
13. Softwareversion<br />
All the factory defaults can be customized in the menu.<br />
All the factory defaults can be restored via the „Factory Default“ option<br />
(with the thermoregulation function switched off). (Also refer to „Salutation“<br />
4.1).<br />
MAIN MENU:<br />
Calibration Prog.<br />
I-> Compressor Auto<br />
Mains Failure Auto<br />
COMPRESSOR AUTO:<br />
I-> Automatic<br />
Always on<br />
Always off<br />
MAIN MENU:<br />
Machine options<br />
I-> Max. heat power<br />
Calibration Prog.<br />
Please enter:<br />
Max. heat power (%)<br />
100<br />
MAIN MENU:<br />
Acoustic alarm<br />
I-> Software version<br />
Set-Point limits<br />
SOFTWARE:<br />
Series No.: 0<br />
Version 04.00s<br />
Date 19.02.03. 15.41<br />
Selection of the compressor starting mode<br />
„Automatic“<br />
The compressor will start automatically<br />
depending on the topical demand. The chiller<br />
will work on demand only.<br />
Benefit: Saving of energy.<br />
Drawback: Heating time (idle time) in the case<br />
of a sudden demand.<br />
„Always On“<br />
The compressor remains switched on,<br />
continuous chiller operation.<br />
„Always Off“<br />
The compressor remains switched of, the<br />
chiller is not in operation.<br />
„Max. Heat Power“ option<br />
Enter the desired maximum heating power in<br />
percent.<br />
This is required for devices equipped with a<br />
heater and a compressor (chiller).<br />
The heating power must be reduced to enable<br />
simultaneous operation of heater and<br />
compressor with a view to the fusing of the<br />
device.<br />
„Software Version“ option<br />
For example:<br />
04.00s vom 19.02.2003, 15.41 PM.<br />
43-4