Calibration<br />
Prog.<br />
4. Thermoregulation via<br />
Controller<br />
4.5. 4.2. Calibration Wichtige Voreinstellungen<br />
prüfen und ändern<br />
1. Calibration Program<br />
1. Programm 52: Rückführung auf Werkseinstellung<br />
2. Programm 90: Sprachauswahl Digitalanzeige UniCop<br />
All the factory defaults can be customized in the menu.<br />
All the factory defaults can be restored via the „Factory Default“ option<br />
(with the thermoregulation function switched off). (Also refer to<br />
„Salutation“ 4.1).<br />
MAIN MENU:<br />
Analog Interface<br />
I-> Calibration prog.<br />
Circulation<br />
2-P.CALIBR. INTERNAL:<br />
I-> Edit T-Cal1/2<br />
Control to T-Cal1<br />
Control to T-Cal2<br />
exit<br />
Please enter:<br />
Set-point TCal1<br />
6 C<br />
Please enter:<br />
Set-point TCal1<br />
10 C<br />
2-P.calibration:<br />
Edit TCal1<br />
I-> Edit TCal2<br />
Control to T-Cal1<br />
Control to T-Cal2<br />
exit<br />
please enter:<br />
set-point TCal2<br />
100 C<br />
Please enter:<br />
set-point TCal2<br />
40 C<br />
2-P.calibration:<br />
Edit TCal1<br />
Edit TCal2<br />
I-> Control to T-Cal1<br />
Control to T-Cal2<br />
exit<br />
2-P.calibration:<br />
Edit TCal1<br />
Edit TCal2<br />
Control to T-Cal1<br />
I-> Control to T-Cal2<br />
Exit<br />
„Calibration Prog.“ option<br />
Is used exclusively to calibrate the internal<br />
sensor.<br />
Use a calibrated reference thermometer as a<br />
second temperature sensor for calibration.<br />
Submenu selection: „Edit TCal1/2“<br />
Entry of the 1 st of two calibration<br />
temperatures, e.g. 10 °C.<br />
Entry of the 2 nd of two calibration<br />
temperatures, e.g. 40 °C.<br />
Submenu selection: „Control to TCal1“<br />
Start thermoregulating until the 1 st calibration<br />
temperature has been reached.<br />
If your reference thermometer indicates the<br />
set temperature reliably, compare the value to<br />
the actual value display of the Polystat<br />
controller. Deviations can be corrected using<br />
the encoder.<br />
Submenu selection: „Control to TCal2“<br />
Proceed as described for TCal1.<br />
Exit the menu once you have completed the<br />
calibration procedure.<br />
The two-point calibration includes all the values between T1 and T2.<br />
With offset calibration, on the other hand, the entire temperature level is<br />
shifted by one value on a linear level.<br />
45-1