4 - HUBER
4 - HUBER
4 - HUBER
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Require<br />
ments<br />
Mechanical<br />
overtemperature<br />
protection<br />
cc1, cc2, cc3,<br />
A5 to A18<br />
B8 to B25,<br />
201,202,<br />
K6 to K25<br />
Electronic<br />
overtemperature<br />
protection<br />
for all units not<br />
listed above as<br />
well as Polystats<br />
with -P04<br />
For detailed information<br />
on the range of<br />
accessories incl. technical<br />
data and price quotations,<br />
please refer to the Huber<br />
catalog or contact your<br />
Huber agent.<br />
2<br />
3<br />
3. Commissioning<br />
3.3. Filling Thermofluid<br />
Overtemperature protection<br />
Prepare the thermostat for<br />
thermoregulation and take safety<br />
measures as described below.<br />
Setting the overtemperature<br />
protection device.<br />
Requirements:<br />
A suitable thermal fluid has been<br />
selected for the process<br />
requirements.<br />
The flash point of the thermal fluid is<br />
known.<br />
For Polystat cc1, cc2, cc3, Polystat with<br />
polycarbonate bath A5 to A18,<br />
Polystat bath thermostat B8 to B25,<br />
heating thermostat 201 and 202,<br />
cooling bath K6 to K25,the over-temperature<br />
protection is set mechanically.<br />
It is independent of the controller.<br />
Use a suitable tool (screw driver or the like)<br />
to turn the red button in the center of the<br />
over-temperature protection device (section<br />
2.2) to the desired temperature value.<br />
33-1<br />
Working temperature range<br />
of thermostat (max.)<br />
Flash point of the thermal<br />
fluid<br />
Over-temperature<br />
protection<br />
(Overheat)<br />
Permissible temperature<br />
range for set-point entry<br />
Procedure:<br />
The over-temperature protection is<br />
set to at least 25 °C below the flash Viscosity limit of the<br />
point of the thermofluid.<br />
thermal fluid<br />
Working temperature range<br />
of thermostat (min.)<br />
Caution!<br />
The overtemperature protection is an especially important safety<br />
device of your thermostat. It should always be operable and be tested<br />
periodically!<br />
2<br />
Mechanical overtemperature<br />
protection<br />
For all other units not listed above, the over-temperature protection<br />
is set electronically. It is independent of the controller.<br />
Use a suitable tool (screw driver or the like) to press the yellow<br />
button in the center of the over-temperature protection device.<br />
The over temperature protection menu will be displayed.<br />
Over temp. protection<br />
OT Setp. heating<br />
OT temp . display<br />
Exit<br />
OT setpoint heating: over temp. value<br />
(adjustable on the Polystat controller)<br />
OT temperature display: display of the<br />
over temperature actual value<br />
3<br />
Electronic overtemperature<br />
protection<br />
Temperature rise