Description of the Custom Features<br />
A customized <strong>Curlin</strong> Infusion <strong>PainSmart</strong> IOD pump may have Maximum Basal Rate<br />
Limit and Maximum Bolus and Dose Limits values that are different from previous default<br />
maximum values for all administrative routes (IV, EPI, and SQ), and a different Lock Level<br />
other than the default value OFF. These values, once programmed at the factory, can not be<br />
changed unless the pump is initialized or a new setting is programmed, both of which must<br />
be performed by <strong>Curlin</strong> or factory trained biomedical professionals.<br />
1 Customized Maximum Basal Rate Limit<br />
A customized <strong>Curlin</strong> Infusion <strong>PainSmart</strong> IOD pump may come with a Maximum Basal<br />
Rate Limit value set in the factory that is different from the default maximum value.<br />
An alert will be displayed if a value greater than the factory set Maximum Basal Rate Limit<br />
value is entered in BASL RATE, <strong>Med</strong>LMTS Mx Rate, and <strong>Med</strong>LMTS Delta Rate.<br />
For example, while programming a PCA (IV) therapy in a customized <strong>Curlin</strong> Infusion<br />
<strong>PainSmart</strong> IOD pump, where the Maximum Basal Rate Limit is factory set to be 9.9 for<br />
IV administration route, if you enter 21 for BASL RATE, an alert message will appear:<br />
“GREATER THAN MAX. FIELD LIMIT MAX=9.9”<br />
The values you are allowed to enter in this example, for BASL RATE, are between 0.1 and<br />
9.9, with an increment of 0.1. Same limit will apply to <strong>Med</strong>LMTS Mx Rate, and <strong>Med</strong>LMTS<br />
Delta Rate also, if the therapy calls for them.<br />