Configuring the GC/MS System for <strong>Pesticide</strong> Analysis<br />

• The BEST PTV Backflush upgrade option for injection volumes < 10 μL inverts the direction of the carrier gas flow in the PTV inlet to vent dirty<br />

matrices and high boilers from the split line. This feature prevents these compounds from entering the analytical column, which significantly reduces<br />

analysis time and extends the life of the column. This option includes a 2 m pre-column.<br />

• For injection volumes greater than 10 μL, the Large Volume Backflush upgrade option allows you to enhance PTV injector performance. This option<br />

enables the solvent venting phase to occur through the backflush system, which allows for a significantly higher recovery of volatiles and makes it possible<br />

for you to increase the injected volume. This option can also be used to eliminate the high boiling part of the sample. Preventing large quantities of<br />

solvent from entering the analytical column and reaching the detector makes large volume PTV easily amenable to mass spectrometers. This option<br />

includes a 10 m pre-column.<br />

The injection parameters for PTV are described in the following tables.<br />