Configuring the GC/MS System for <strong>Pesticide</strong> Analysis<br />
TSQ Quantum GC Mass Spectrometer<br />
The TSQ Quantum GC mass spectrometer ships with the recommended configuration for pesticide analysis. It is equipped for easy ion volume exchange.<br />
The method settings for the TSQ Quantum GC will provide good results for all listed compounds. Adjustments can be made if the compound list changes or<br />
additional sensitivity is required.<br />
Table 22. TSQ Quantum GC Mass Spectrometer - Tuning Parameters<br />
AutoTune<br />
Setting<br />
Parameter<br />
Manual tune may be recommended for special mass range enhancements.<br />
Table 23. TSQ Quantum GC Mass Spectrometer - Source Parameters<br />
Setting<br />
Ionization Mode EI<br />
Emission Current 50 uA<br />
Ion Volume<br />
Closed EI Volume<br />
Source Temperature 250 °C<br />
Parameter<br />
Tip Even with a maximum GC oven temperature of 300 °C, a low transfer line temperature works well and prevents active<br />
sites created in a constantly high heated transfer line.<br />
44 <strong>Pesticide</strong> <strong>Analyzer</strong> <strong>Reference</strong> Thermo Scientific