Configuring the GC/MS System for <strong>Pesticide</strong> Analysis<br />

Table 18. TriPlus AS or AS 3000 II Autosampler - Sampling Parameters<br />

Setting<br />

Parameter<br />

Sample Volume 1.0 - 2.0 μL. PTV/SSL large volume (LVI) conditions require different oven, injector temperature program and timing settings.<br />

Refer to the TRACE GC Ultra documentation for details.<br />

Plunger Strokes 5<br />

Air Volume<br />

1.0 μL<br />

Filling Volume<br />

2.0 μL<br />

Table 19. TriPlus AS or AS 3000 II Autosampler - Vial Sampling Depth Parameters<br />

Vial Depth<br />

Setting<br />

Bottom<br />

Parameter<br />

Table 20. TriPlus AS or AS 3000 II Autosampler - Injection Mode Parameters<br />

Setting<br />

Pre-injection Dwell Time<br />

Post-injection Dwell Time<br />

0.0 s<br />

2.0 s<br />

Parameter<br />

Table 21. TriPlus AS or AS 3000 II Autosampler - Viscosity Parameters<br />

Sample Type<br />

Setting<br />

Non-viscous<br />

Parameter<br />

Thermo Scientific <strong>Pesticide</strong> <strong>Analyzer</strong> <strong>Reference</strong> 43