Configuring the GC/MS System for <strong>Pesticide</strong> Analysis<br />

SSL Injector<br />

There are two configurations of the SSL injector for pesticide analysis:<br />

• The Optimized Geometry SSL injector upgrade option is used for injecting 1-5 μL samples at a constant temperature.<br />

Note Labile pesticides, as indicated in the list of compound classes, cannot be analyzed at optimum performance. For analysis of those compounds, the<br />

B.E.S.T. PTV injector is recommended.<br />

• The Large Volume SSL injector upgrade option allows you to introduce volumes as high as 50 μL in a standard split-splitless injector. Using Concurrent<br />

Solvent Recondensation (CSR, a Thermo Fisher Scientific patented technology) enables you to overcome the limitations of the liner’s internal volume.<br />

The solvent vapors are recondensed in the pre-column and transferred to the analytical column. CSR large volume injection can be performed manually<br />

or automatically with the AI/AS 3000 II and TriPlus autosamplers.<br />

The injection parameters for SSL are described in the following table.<br />

36 <strong>Pesticide</strong> <strong>Analyzer</strong> <strong>Reference</strong> Thermo Scientific