Configuring the GC/MS System for <strong>Pesticide</strong> Analysis<br />
• Cooled/heated trays for TriPlus:<br />
−<br />
−<br />
Primary cooled/heated tray with 1-96 labeling for 1, 2, 2.5 mL vials<br />
Secondary cooled/heated tray with 97-192 labeling for 1, 2, 2.5 mL vials<br />
• 2x100 mL washing station<br />
• 100 μl syringe (0.63/50/side hole) with removable needle suitable for PTV-LVI (PN 365 20050)<br />
• 100 μl syringe (0.63/50/cone) with removable needle suitable for LV-SL (PN 365 00495)<br />
The autosampler parameters are described in the following tables.<br />
Table 16. TriPlus AS or AS 3000 II Autosampler - Syringe Parameters<br />
Setting<br />
Needle Length<br />
80 mm for SSL<br />
50 mm for PTV<br />
Parameter<br />
Table 17. TriPlus AS or AS 3000 II Autosampler - Mode, Injector, Sync Parameters<br />
Setting<br />
Injection Mode Basic<br />
Injector Port Injector A (or B)<br />
Start Sync Mode Standard<br />
Parameter<br />
42 <strong>Pesticide</strong> <strong>Analyzer</strong> <strong>Reference</strong> Thermo Scientific