23.03.2015 Views

Pesticide Analyzer Reference - writeframeofmind.biz

Pesticide Analyzer Reference - writeframeofmind.biz

Pesticide Analyzer Reference - writeframeofmind.biz

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Configuring the GC/MS System for <strong>Pesticide</strong> Analysis<br />

Table 5.<br />

PTV Injector - QC Sample Parameters<br />

Compound<br />

Setting<br />

Endrin<br />

Parameter<br />

QA/QC check: The breakdown products Endrin-aldehyde and Endrin-ketone must be < 20% of Endrin peak area. If<br />

> 20%, maintenance is recommended.<br />

Table 6.<br />

PTV Injector - Solvent Parameters<br />

Setting<br />

Solvents Cyclo-hexane/Acetone 9:1<br />

Parameter<br />

• Beware of polar solvents that moisten the column film.<br />

• Do not use 100% acetone.<br />

• Acetonitrile (ACN) is not suitable for GC injection (only LC) because it strips the deactivation and damages<br />

the column film.<br />

• With QuEChERS sample preparation, solvent exchange is necessary, so small amounts of ACN in a suitable<br />

solvent are acceptable.<br />

• Alternative solvents are ethylacetate and ethylacetate/cyclohexane at 1:1 mix (or Ethyl Acetate/Cyclohexane 1:1)<br />

Thermo Scientific <strong>Pesticide</strong> <strong>Analyzer</strong> <strong>Reference</strong> 33