10.07.2015 Views

IQ plus® 800/810 - Rice Lake Weighing Systems

IQ plus® 800/810 - Rice Lake Weighing Systems

IQ plus® 800/810 - Rice Lake Weighing Systems

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

3.2.8 Analog Output MenuSee Section 9.11 on page 73 for information about installing, configuring, and calibrating the analog outputoption.CONFIGXXXXXXX SET ALGFORMATSETPNTSSERIALP XXXXXXX FORMTDIG INALG XXXXXXX OUT BAR XXXXXXX GRF CALIBRT VERSION XXXXXXXALGOUT1ALGOUT2SOURCEZEROSPANsame as ALGOUT1TOT GnumbernumberTOT NCH1 GCH1 NCH2 GCH2 NCH3 GCH3 NCH4 GCH4 NFigure 3-12. Analog Output MenuALG OUT MenuParameter Choices DescriptionLevel 2 submenusALGOUT1ALGOUT2SOURCEZEROSPANLevel 3 submenusSOURCE TOT GTOT NCH1 GCH1 NCH2 GCH2 NCH3 GCH3 NCH4 GCH4 NSpecifies the level 3 parameter being set for Analog Output 1 or 2.Specifies the source channel tracked by the analog output. Either gross or net values can beselected.ZERO number Zero calibration value. Use keypad to enter zero value. See page 75 for calibration instructions.SPAN number Span calibration value. Use keypad to enter span value. See page 75 for calibrationinstructions.Table 3-8. Analog Output Menu ParametersConfiguration 35

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!