HPLC-8100 HPLC Chromatograph2. Rear view of UV detectorFig. 11 Rear view of UV detector1.RS—232 interface3. Power supply, ON/OFF, and self-test2.Remote terminal3. Analog output 4. 24V ventilator5.12V ventilator 6. Power switch7., power socket 8. Slow melting fuse(1A)UV detector operates with 220±20V AC of 50±0.5Hz. 1A slow melting fuse is used. Thesetting is done by the manufacture on custom request. The actual setting is indicated onrear panel of the instrument (Fig. 12).theCaution:Cut off power supply before opening cover.220±20V 50±0.5 Hz100VAFUSE: 1A SLOWFig. 12 Indication of the input voltageCAUTION: Make certain that the correct voltage has been set on the rear panel of theinstrument, the power supply is reliably grounded and a corresponding 3-pole power cableis used.Connect the detector to the power supply. Switch on the instrument. Power switch is on the rearpart of the instrument.After switching on, the instrument first performs the self-test as follows:BFRL UV/VIS DETECTORRAM TEST OK14