EnteraLite Operator's Manual - Moog Inc
EnteraLite Operator's Manual - Moog Inc
EnteraLite Operator's Manual - Moog Inc
- No tags were found...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
CHAPTER 2 - DIRECTIONS FOR USELoading the Disposable SetCAUTION: Never press fingers or other objects into the cavities over thepressure sensors. Load only the silicone tubing of the disposable set intothese sensors.CAUTION: Dispose of <strong>EnteraLite</strong> ® Sets Properly, as required by local law.To load the disposable set, startwith the tube fitting closest tothe feed bag. Position theround fitting on the outside ofthe pump and then press thesilicone tubing into the air sensoras shown in Figure 2-9.Figure 2-9: Load Tubing in Air SensorNext, route the silicone tubingsegment through the upstreampressure sensor area andtoward the pump wheel asshown in Figure 2-10. Do notstretch the tubing at this point.Figure 2-10: Load Tubing in Upstream Pressure Sensor8