CHAPTER 2 - DIRECTIONS FOR USEAn Interval Feeding ExampleSuppose you want to deliver 300 ml of enteral feeding solution at a rate of 80 ml/hr, andrepeat this feeding every 6 hours:Turn the pump on by pressing the ON/OFFkey. Verify that all display segments and symbolsare active as shown in Figure 2-4, andthat the audible alarm sounds. Once thepump finishes its self-test the rate will be displayedby the large characters.Press the MORE key or the LESS key toincrease or decrease the rate to 80 ml/hr.Figure 2-4: Display SegmentsPress the RATE/DOSE key to display the dose. Press the MORE or LESS key until adose of 300 ml is displayed.Now press the FEED INT key. Press the MORE or LESS key until the feed interval isset to 6 hr.WARNING: Replace the disposable set every 24 hours to maintain deliveryaccuracy and prevent growth of harmful bacteria.WARNING: Low flow rates, combined with high dose settings may exceedthe life of the disposable set. The disposable set must be replaced every 24hours to maintain delivery accuracy and prevent the growth of bacteria. DONOT PROGRAM RATE AND DOSE COMBINATIONS WHICH EXCEED A 24HOUR FEED REGIMEN.11