CHAPTER 2 - DIRECTIONS FOR USEAn Interval Feeding ExampleOnce the feed interval, dose, and ratehave been set and the RUN/PAUSE key ispressed, the pump will keep running untilthe bag or solution container is empty.Between interval feed cycles, the wordsrun and FEED INTERVAL will blink, butthe arcs around run will not appear. Also,the large characters will display the messageNEXT DOSE ##.## hr, where theformat ##.## indicates the time inhours.minutes until the next dose. Anexample is shown in Figure 2-13 (3.59 hrequals three hours and fifty-nine minutes).Figure 2-13: Time Displayed BetweenInterval FeedingsNOTE:On early versions of the pump, the next dose timer is displayed using the format##.# hr (As an example 3.9 hours equals 3 hours and 54 minutes. Every 0.1 hrequals 6 minutes).13