2.3 Connections<br />

<strong>Q150T</strong> <strong>Sample</strong> <strong>Preparation</strong> <strong>System</strong><br />

Installation consists of the following steps:<br />

� Connect gas supplies (see below).<br />

� Make electrical connections (see page 12).<br />

� Connect the vacuum pump (see page 13).<br />

� Install the sample stage (see page 14).<br />

If you want to install a different insert than that supplied with your instrument, please see<br />

page 49. If you have ordered a <strong>Q150T</strong> <strong>ES</strong>, the instrument will be fitted with the sputter<br />

insert, unless you have specifically requested another configuration.<br />

Figure 2-1 shows the location of connectors and switches on the rear panel of the<br />

<strong>Q150T</strong>.<br />

POWER INLET PUMP OUTLET AUX GAS INLETS VACUUM CONNECTION<br />

Power inlet with<br />

integral On/Off<br />

rocker switch<br />

2.3.1 Gas Connections<br />

Power supply for rotary<br />

pump allowing control by<br />

instrument.<br />

Figure 2-1. <strong>Q150T</strong> Rear Panel<br />

Network<br />

connection<br />

Argon<br />

0.3bar<br />

(4psi)<br />

Nitrogen<br />

0.3bar<br />

(4psi)<br />

Rotary Pump Vacuum<br />

connection<br />

WARNING<br />

UNDER NO CIRCUMSTANC<strong>ES</strong> SHOULD ANY OTHER<br />

CONNECTIONS OR OUTLETS/INLETS BE USED FOR ANY OTHER<br />

EQUIPMENT OR SERVIC<strong>ES</strong>.<br />

1. Connect a suitable argon supply at a regulated input of 0.3bar (4psi) to the quick<br />

connect Argon gas connector on the rear panel (see Figure 2-1.<br />

2. Connect a suitable nitrogen supply at a regulated input of 0.3bar (4psi) to the<br />

quick connect Nitrogen connector on the rear panel. We do NOT recommend<br />

using air as the purge gas.<br />

If you prefer, you can use argon for both process and purge supplies. To do so, install a<br />

'T' piece across both gas inlets.<br />

Connectors are push-fit and will 'snap' into a locked position. To release a connector,<br />

depress the metal tongue.<br />

<strong>Q150T</strong> - <strong>Instruction</strong> <strong>Manual</strong> 11 10473 - Issue 3