7.11 Extended Height Cylinder<br />

<strong>Q150T</strong> <strong>Sample</strong> <strong>Preparation</strong> <strong>System</strong><br />

1. Check the machine is not operating.<br />

2. Press down firmly on the top cover to release the latch.<br />

3. Lift the top cover and rest it against its stop.<br />

4. Remove the standard glass cylinder assembly (10068).<br />

5. Remove the M6 Cap head screw from the front of the pillar, slide the whole top<br />

cover and inner pillar assembly upwards until you see the fourth hole line up and<br />

replace the M6 cap head screw.<br />

6. Place the extended glass cylinder assembly 10429 in place and close the top<br />

cover.<br />

The vent times will need to be extended to approx 200 seconds in profiles<br />

using the extended height cylinder.<br />

Figure 7-17. The extended height cylinder in position<br />

7.12 Full Range Vacuum Gauge Assembly<br />

The instrument must be returned to the factory for fitting this option as the<br />

covers need to be removed.<br />

As standard, the <strong>Q150T</strong> is fitted with a Pirani gauge with a measurement range of 1000<br />

– 5x10 -4 mbar. The typical ultimate vacuum of the pumping system in a clean instrument<br />

after pre-pumping with dry nitrogen gas is 5 x10 -5 mbar. When the vacuum is better than<br />

5x10 -4 mbar the instrument simply displays “HiVac”. For users who need an accurate<br />

value for the vacuum achieved before and during the coating process (for example,<br />

when evaporating carbon it is desirable to achieve as good a vacuum as possible), the<br />

Full Range Vacuum Gauge Assembly option (10428) extends the accurate measurement<br />

range to 1000 – 5x10 -9 mbar.<br />

Contact Quorum for further information.<br />