The above page allows you to calibrate the low point. Please adjust your input whichconnects to the channel terminal strip to a low point value closes enough to the lowest valueof the channel. In the above example “0” is the lowest value.Click Start Calibration button to start the low point calibration. After the click, StartCalibration button changed to Stop Calibration.During the calibration process the current real-time value is showing in Current Readingfield. If you think this value is stable you may click Stop Calibration button.Click Next to proceed to the next page.<strong>SiteView</strong> User’s <strong>Manual</strong> – Microedge Instruments Inc. 57/200