Manual capaNCDT 6300 / 6310 (PDF, 1.74 MB) - Micro-Epsilon
Manual capaNCDT 6300 / 6310 (PDF, 1.74 MB) - Micro-Epsilon
Manual capaNCDT 6300 / 6310 (PDF, 1.74 MB) - Micro-Epsilon
- No tags were found...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Functional Principle, Technical Data2.3 Technical DataController typeResolution staticResolution dynamicLimit frequencyLimit frequency adjustableLinearityMax. sensitivity deviationLong term stabilitySynchronous operationInsulator measurementFSO = Full Scale Output<strong>capaNCDT</strong> <strong>6300</strong> / <strong>6310</strong>DT<strong>6300</strong>/DT<strong>6310</strong>0.001 % FSO0.01 % FSO (8 kHz)8 kHz20 Hz / 1 kHz / 8 kHz±0.2 % FSO (all sensors interchangeable without calibration)Option LC: ±0.1 % FSO (tuned to one sensor)±0.1 % FSO≤ 0.02 % FSO / monthTemperature stability ±0.01 % FSO / °CTemperature range (operation) +10 … +50 °CTemperature range (storage) -10 … +75 °CSupplyOutputSuitable for sensorsSensor cable standardSensor cable (matched)Proper environment sensorProtection classyesyes±15 VDC (±2 %) / ±150 mA0 - 10 VDC (max. 10 mA short circuit proof)4 ... 20 mA (load max. 500 Ω)all sensors≤ 1 mup to 4 mHumidity 5 to 95 % (non condensing)IP 54 (Controller and sensors)Electromagnetic compatibility (EMC) DIN EN 61326-1: 2006-10 and DIN EN 61326-2-3: 2007-5Page 15