Functional Principle, Technical Data2.3 Technical DataController typeResolution staticResolution dynamicLimit frequencyLimit frequency adjustableLinearityMax. sensitivity deviationLong term stabilitySynchronous operationInsulator measurementFSO = Full Scale Output<strong>capaNCDT</strong> <strong>6300</strong> / <strong>6310</strong>DT<strong>6300</strong>/DT<strong>6310</strong>0.001 % FSO0.01 % FSO (8 kHz)8 kHz20 Hz / 1 kHz / 8 kHz±0.2 % FSO (all sensors interchangeable without calibration)Option LC: ±0.1 % FSO (tuned to one sensor)±0.1 % FSO≤ 0.02 % FSO / monthTemperature stability ±0.01 % FSO / °CTemperature range (operation) +10 … +50 °CTemperature range (storage) -10 … +75 °CSupplyOutputSuitable for sensorsSensor cable standardSensor cable (matched)Proper environment sensorProtection classyesyes±15 VDC (±2 %) / ±150 mA0 - 10 VDC (max. 10 mA short circuit proof)4 ... 20 mA (load max. 500 Ω)all sensors≤ 1 mup to 4 mHumidity 5 to 95 % (non condensing)IP 54 (Controller and sensors)Electromagnetic compatibility (EMC) DIN EN 61326-1: 2006-10 and DIN EN 61326-2-3: 2007-5Page 15