EiUK USER MANUAL RASI 700<br />

12.2 Core flow search<br />

Before using the core flow search it must be switched-ON:<br />

OK.<br />

The core flow search will help you to find the optimal measurement point<br />

in the stack. The core flow can be identified by the maximum flue gas<br />

temperature.<br />

In high reaction time the analyzer displays the trend of the flue gas<br />

temperature. Insert the probe pipe slowly into the stack and position<br />

your probe tube when you have reached the maximum flue gas temperature<br />

that is displayed.<br />

Approaching<br />

the maximal flue gas temperature<br />

Temperature rising<br />

Positioning the probe in the core flow:<br />

Insert the probe pipe slowly into the stack and position your probe pipe<br />

when you have reached the maximum flue gas temperature that is displayed<br />

(see temperature maximum value on the display – in this case 69°C).<br />

Maximum temperature has been reached when the arrows (left picture) disappear,<br />

max. (right picture) appears in place of the arrow, and the<br />

beeper signal stops. Moving away from the max. temperature will result in<br />

the bars moving away from the max. temperature (1 bar is equivalent to<br />

1°C). Once the right core flow has been achieved, the probe is fixed with<br />

the probe cone screw.<br />