EiUK USER MANUAL RASI 700<br />
14.5 Leak proof test<br />
With the leak proof test the system is checked by the device (incl. the<br />
condensate separator) up to the probe spike on undensity. The internal<br />
gas pump generates in addition a subpressure which is measured over the<br />
built-in draft sensor and is observed for a period of 10 seconds. Based<br />
on the decrease of pressure the leakness of the system will be determined.<br />
Operation:<br />
• The leak proof test cap # 61382 (for probe tubes Ø 8 mm) must be<br />
put on on the probe spike.<br />
ATTENTION:<br />
The probe spike before the density test clean!<br />
(With depositions on the tube the cap does not<br />
seal.)<br />
• Launch under “extras” the leak proof test on which the following<br />
window will pop up:<br />
OK<br />
…<br />
If of the leak proof test is not passed the probe must be checked including<br />
the hosing as well as the condensate separator.<br />