FVA/A1 STERILIZERS
FVA/A1 STERILIZERS
FVA/A1 STERILIZERS
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>FVA</strong>/<strong>A1</strong> - OPERATION AND<br />
MAINTENANCE MANUAL<br />
When electric power returns, an acoustic alarm is issued and the operator must recognize the emergency state by acting<br />
on the mute icon of the touch-screen;<br />
The operator must then leave the emergency state by acting on the STOP icon of the touch-screen;<br />
The sterilizer is thus brought to the atmospheric pressure balancing phase.<br />
If the sterilizer was not in a phase of an operating cycle:<br />
when electric power returns, the Fedegari logo is displayed while waiting for commands from the operator.<br />
The sterilization chamber is provided with a pneumatic gasket for sealing the lid and with associated devices for ensuring the<br />
tightness of the chamber even if there is no electric or pneumatic power.<br />
In particular, depressurization of the pneumatic gasket of the door takes longer than the depressurization of the chamber.<br />
Atmospheric pressure in the chamber, caused by failure of the tightness of the gasket as a consequence of a blackout, is reached<br />
after at least 20 hours, depending on the overall maintenance level of the machine.<br />
2.8.2. FAILURE OF WATER SUPPLY<br />
In order to ensure that the steam generator is supplied with pressurized water at 1 to 4 bars, the sterilizer is provided with a selfpriming<br />
pump, which is capable of drawing water from any container located outside the sterilizer or from the water supply mains.<br />
In order to protect its integrity, the self-priming pump is equipped with a suitable internal device for checking the minimum level of the<br />
liquid.<br />
Any interruption in the water supply of the steam generator increases the time required to complete the process phases in progress. If<br />
the water supply fails, the sterilizer issues an acoustic alarm signal and is then forced by the operator into a state known as<br />
“emergency” and waits for the optimum parameters to be stored.<br />
� NOTE<br />
If the optional Vacuum and drying assembly kit or the Quick drying assembly kit is installed, the sterilizer is<br />
equipped with a vacuum pump and/or with an internal cooling plate.<br />
Interruption of the water supply of the vacuum pump and/or of the internal cooling plate increases the time required<br />
to complete the phases of the process in progress. In this case, it is possible to force the sterilizer into a state<br />
known as “emergency” while waiting for the optimum parameters to be restored.<br />
2.8.3. SHUTDOWN<br />
The machine is equipped with a main circuit breaker, which is located on the rear panel. Opening this circuit breaker completely stops<br />
the supply of electric power to the machine.<br />
The main circuit breaker disconnects all the poles of the power supply line.<br />
2.8.4. EMERGENCY STOP<br />
The machine is provided with an emergency switch, which is located in the lower part of the post of the process controller. Activation<br />
of the mushroom-type emergency pushbutton, provided with a mechanical reset, disconnects the electric power supply of the<br />
sterilizer.<br />
If the sterilizer was in an operating cycle phase:<br />
When electric power returns, an acoustic alarm signal is issued and the operator must acknowledge the state of<br />
emergency by acting on the mute icon of the touch-screen;<br />
Tthe operator must then leave the state of emergency by acting on the STOP icon of the touch-screen. The sterilizer is<br />
thus brought to the atmospheric pressure balancing phase.<br />
If the sterilizer was not in an operating cycle phase:<br />
when electric power returns, the Fedegari logo is displayed while waiting for commands on the part of the operator.<br />
ID no. 85693 Rev. 2 10 / 31 July 2005