FVA/A1 STERILIZERS
FVA/A1 STERILIZERS
FVA/A1 STERILIZERS
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>FVA</strong>/<strong>A1</strong> - OPERATION AND<br />
MAINTENANCE MANUAL<br />
- if the optional function (safety temperature) is set, the heat probe arranged in the chamber must detect a temperature that does<br />
not exceed the set value, making sure that any liquids placed to be sterilized in glass containers inside the autoclave have<br />
reached a temperature that is lower than their boiling point at atmospheric pressure;<br />
- once the process controller has enabled lid opening, the air introduced in the gasket can be released if the residual pressure in<br />
the chamber has been checked to be lower than 50 mbar by means of a pressure-controlled switch. This check is performed with<br />
fail-safe criteria.<br />
Release of the air fed into the gasket seat<br />
Once the clearances listed in the preceding subsection have been obtained, if the operator presses the door release button, the<br />
compressed air is discharged from the gasket seat.<br />
Manual opening of the autoclave<br />
After automatic discharge of the air from the sealing gasket seat, the opening maneuver is completed manually.<br />
3.7.2. SAFETY VALVES<br />
To ensure total safety of the operator, the sterilizer is equipped with the following devices:<br />
� a safety valve that discharges the excess pressure inside the chamber;<br />
� a safety valve that discharges the excess pressure in the autonomous steam generator.<br />
� NOTE<br />
Please refer to the technical manual of the machine for a detailed description of the functions and checks to be<br />
performed on the safety valves.<br />
3.8. TECHNICAL DATA<br />
3.8.1. MAIN DIMENSIONS AND WEIGHTS<br />
The main dimensions of the vertical steam sterilizer related to the <strong>FVA</strong>/<strong>A1</strong> range are listed in the following table:<br />
Model<br />
Main data of the various models<br />
<strong>FVA</strong>2/<strong>A1</strong> <strong>FVA</strong>3/<strong>A1</strong><br />
Useful loading capacity [l] 75 140<br />
Chamber type vertical with cylindrical cross-section, top-loading<br />
Chamber diameter [mm] 400 500<br />
Useful height [mm] 600 700<br />
Weight when empty [kg] 150 175<br />
Weight when filled with water [kg] 270 340<br />
Loading height [mm] 875 875<br />
External dimensions (1) W x H x D [mm] 1200 × 1305 × 600 1200 × 1305 × 600<br />
Notes to the table:<br />
(1) External dimensions measured when the lid is closed.<br />
3.8.2. TECHNICAL PARAMETERS<br />
� NOTE<br />
Please refer to the technical manual of the machine for a detailed description of the functional, electrical and<br />
construction parameters.<br />
ID no. 85693 Rev. 2 14 / 31 July 2005