5.4.1. HYDRAULIC CONNECTIONS<br />
� NOTE<br />
<strong>FVA</strong>/<strong>A1</strong> - OPERATION AND<br />
MAINTENANCE MANUAL<br />
Please refer to the technical manual of the machine for a detailed description of the hydraulic connections.<br />
5.4.2. CONNECTIONS TO THE ELECTRIC POWER SUPPLY<br />
Connection of the power supply cable of the machine to the electrical mains is under the responsibility and care of the customer.<br />
When performing the connection, strict compliance with the following is recommended:<br />
! WARNING!<br />
Connection to the electrical mains must be performed exclusively by qualified personnel, fully complying with all the<br />
safety requirements prescribed by applicable standards with reference to the installation site of the machine.<br />
All connections must be made with no mains voltage present.<br />
! WARNING!!<br />
Please refer to the technical manual of the machine in order to determine the correct polarities and connection<br />
sequence.<br />
Connecting the ground conductor poorly, or not connecting it at all, may seriously compromise the operating safety<br />
of the sterilizer. FEDEGARI declines any responsibility arising from connecting the ground conductor poorly or not<br />
at all.<br />
� NOTE<br />
Please refer to the technical manual of the machine for a detailed description of the electrical connections.<br />
5.5. CHECKING THE INSTALLATION<br />
Before normal use of the sterilizer, it is necessary to provide a short running-in period, which consists in performing at least two<br />
complete cycles while the machine is empty.<br />
During this period, it is necessary to monitor and search for any anomalies.<br />
The machine can be introduced in the normal production cycle only after these tests have been successful.<br />
If anomalies in operation are observed, contact the manufacturer immediately.<br />
� NOTE<br />
Please refer to the technical manual of the machine for further information on the installation of the sterilizer.<br />
ID no. 85693 Rev. 2 18 / 31 July 2005