4. TRANSPORT, HANDLING AND STORAGE<br />
<strong>FVA</strong>/<strong>A1</strong> - OPERATION AND<br />
MAINTENANCE MANUAL<br />
4.1. TRANSPORT<br />
If the sterilizer is to be transported outside the destination areas, it is recommended to follow the procedures for the loading and<br />
unloading operations described in the technical manual of the machine.<br />
! WARNING!<br />
Improper use of the lifting or anchoring devices of the machine can cause severe damage to the machine and to its<br />
operators.<br />
4.2. HANDLING<br />
All <strong>FVA</strong>/<strong>A1</strong> models are provided with a suitable stand on a set of four wheels, which ensures easy handling both during installation<br />
and during maintenance or emergency.<br />
Final placement of the machine at the installation site is generally performed by two operators.<br />
To ensure correct and safe handling of the machine, it is recommended to perform the following steps:<br />
1. remove the enclosure that protects the machine, separating it from the platform that constitutes the base of the package;<br />
2. transport the machine, keeping it fixed to the base of the package, to the installation site, and then deposit it on the floor of the<br />
room;<br />
3. remove the locking devices;<br />
4. lift the machine, freeing it from the pallet.<br />
! WARNING!<br />
If the handling of the machine requires lifting, check that the cables and the lifting means are suitable for the<br />
required capacities and have the approvals required by the relevant local laws.<br />
4.3. STORAGE<br />
The machine can be stored in a closed dry room with a temperature comprised between 4°C and 50°C and a relative humidity of no<br />
more than 80%.<br />
If rooms suitable to ensure these environmental conditions for machine storage are not available, it is advisable to ask FEDEGARI to<br />
supply the machine with a heat-sealed enclosure made of multi-ply barrier material in a sea case.<br />
! WARNING!!<br />
Sunlight can cause severe damage to the plastic and rubber parts. Direct exposure to the sun for prolonged periods<br />
must be avoided.<br />
; CAUTION<br />
Storage periods of less than two months are recommended. If longer periods are necessary, the gaskets should be<br />
treated with silicone oil in order to prevent early deterioration or unwanted sticking. After particularly long storage<br />
periods, it may be necessary to replace gaskets and seals before returning the machine to production status.<br />
ID no. 85693 Rev. 2 16 / 31 July 2005