6.4.3. PRODUCT LOADING<br />

� Open the lid of the chamber by turning it counterclockwise;<br />

� Remove the temperature monitoring sensor from the sterilization chamber<br />

� Use the appropriately provided devices and accessories to load the product to be sterilized<br />

� Place the Pt100 flexible temperature control sensor and any optional component listed below:<br />

� optional heat probe;<br />

� safety thermal block device.<br />

in the upper drum so that their metal tips are in contact with the product.<br />

� NOTE<br />

During product loading, pay particular attention to avoid damaging the heat probes.<br />

<strong>FVA</strong>/<strong>A1</strong> - OPERATION AND<br />

MAINTENANCE MANUAL<br />

In order to facilitate the loading of the product to be sterilized, it is recommended to use the appropriately provided drums for<br />

positioning the load. These drums are made of AISI 304 stainless steel, are provided with a handle to allow easier transport or<br />

coupling to the optional electric winch, and have a stainless steel supporting plate. The structure is a perforated mesh, which is meant<br />

to avoid retaining condensed water at the end of the sterilization process. The shape of the drum is designed in order to contain the<br />

various loads to be sterilized that are normally used in laboratories.<br />

Figure 6.4.3.a- Loading operation Figure 6.4.3.b- Loading devices<br />

� NOTE<br />

For further information regarding the various available drum models of drum, please refer to Section 1 of the<br />

TECHNICAL MANUAL.<br />

! CAUTION!<br />

During product loading operations, check that the load is arranged evenly inside the drum in order to avoid<br />

imbalances during lifting.<br />

If the final weight of the drums makes them difficult to handle due to the loading of the product in the drums, it is advisable to use the<br />

appropriately provided load lifting device, which can be supplied as an optional kit.<br />

IThe load lifting device falls within the scope of workplace safety standards and is constituted by an electrically-driven winch (Fig.<br />

6.4.3.c) and by the associated coupling accessory (Fig. 6.4.3.d). The horizontal movement of the oscillating arm of the winch is<br />

manual, while the vertical movement is provided by an electric motor. For safety reasons, the movement of the arm of the winch is<br />

interrupted when the control pushbutton is released. The winch structure and the arm are made of AISI 304 stainless steel.<br />

ID no. 85693 Rev. 2 22 / 31 July 2005